An Andersen sampler, Andersen impactor, or sieve impactor[2] is a cascade impactor used to determine the amount of viable pathogens in a given area, in particular bacteria and fungi. Unlike real-time electronic particle counters, the Andersen sampler imparts pathogens on petri dishes, which require incubation. Thus, calculation of the contaminated air requires working backwards from the resulting pathogen growth in each dish.
Andersen samplers have been used to assess the nature of pathogenic aerosols in various scientific papers. It can also be used to study inert aerosols, when each stage of the sampler is used to model respiratory particle deposition.
Operation
Andersen's paper from 1958 describes a six-stage Andersen sampler that counts "viable airborne particles". In each stage, air goes through specifically sized holes, past a petri dish that aerosols impact, before subsequently moving through smaller holes towards the other petri dishes. Each stage gradually increases the velocity of the air; the sampler relies on inertia for particles to leave the air stream.[1]

Calculation of the number of viable particles is normally done by counting colonies via a microscope, called the "microscope method" in Andersen's paper. For stages 3 through 6, Andersen provides an alternate means of calculating viable particles via the "positive hole method". The method involves counting the number of colonies from a macroscopic point of view; each visible colony corresponds to a hole in each stage of the Andersen sampler. A conversion table is then used to quickly calculate the number of viable particles.[1]
Los muestreadores Andersen se han utilizado en diversos lugares e industrias, incluyendo (pero no limitándose a) instalaciones de compostaje, para ayudar a mitigar los impactos en la salud de las esporas de hongos, [ 3 ] agricultura, para rastrear bioaerosoles de manera más general provenientes del ganado , [ 4 ] textiles, [ 5 ] edificios de apartamentos, [ 6 ] instalaciones de aguas residuales, [ 7 ] e incluso la aerosolización de ántrax después de los ataques en 2001. [ 8 ] Los muestreadores también se han utilizado para estudiar el polvo inerte de plomo, con cada etapa diseñada para simular las características de deposición del tracto respiratorio. [ 9 ]
Historia
La información sobre el desarrollo del muestreador Andersen fue clasificada inicialmente por el ejército estadounidense , pero se desclasificó el 24 de enero de 1958. [ 10 ] Un uso temprano destacado del muestreador Andersen fue el seguimiento de enfermedades en los estados de Oregón y California. [ 11 ]
El desarrollo continuo del muestreador Andersen fue entonces el objetivo principal de Andersen 2000, Inc., originalmente CMC Industries, que cambió su nombre tras la adquisición de las patentes correspondientes. En noviembre de 1970, la compañía había lanzado cuatro modelos del muestreador Andersen, cada uno dirigido a diferentes industrias, incluyendo uno portátil. Sin embargo, el muestreador de apilamiento continuó siendo el modelo más popular de la compañía. [ 12 ]
En 1988 se publicó una evaluación del impactador en cascada Andersen Mark-II. [ 13 ]
En 2012, se realizó una comparación entre el método de "partículas cultivables" (CP) y el método de "organismos cultivables" (CO) con la ayuda de un analizador aerodinámico de partículas (APS). Las distribuciones entre los métodos CP y CO fueron lo suficientemente diferentes como para rechazar la hipótesis nula; los autores sugieren que esto puede deberse a la mayor resolución en el recuento de partículas que ofrece el APS, a las pérdidas en las paredes del muestreador Andersen o a la recolección involuntaria de partículas más pequeñas en las etapas iniciales. [ 14 ]
En 2013, un grupo de investigadores mejoró la eficiencia de recolección del muestreador de Andersen agregando aceite mineral al agar de la placa de Petri. Los resultados fueron respaldados mediante el uso de un contador óptico de partículas. [ 15 ]
Véase también
- Impactador en cascada : método gradual de medición y recolección de aerosoles.
- Impacto de aerosoles
- Muestreador de partículas
- Bioaerosol
- Contador de partículas
Referencias
- 1 2 3 Andersen, Ariel A. (1958). "Nuevo muestreador para la recolección, dimensionamiento y enumeración de partículas viables en el aire" . Journal of Bacteriology . 76 (5): 471– 484. doi : 10.1128/jb.76.5.471-484.1958 . PMC 290224. PMID 13598704 .
- ↑ "Muestreo de aire". DIRECTRICES PARA LA EVALUACIÓN DE BIOAEROSOLES EN AMBIENTES INTERIORES . Cincinnati, Ohio: Conferencia Americana de Higienistas Industriales Gubernamentales. 1989. ISBN 0-936712-83-X.
- ↑ Feldman, K. (1995). "Muestreo de contaminantes en el aire" . Biocycle . 36 (8): 84. ProQuest 236870502 .
- ↑ Cole, NA; PAS, Todd R.; Auvermann, B.; Parker, D. (2008). "Auditoría y evaluación de la calidad del aire en operaciones de alimentación concentrada1,2,3" . Professional Animal Scientist . 24 (1): 1– 22. doi : 10.15232/S1080-7446(15)30804-4 . ProQuest 230415564 .
- ↑ Rubbo, Sydney D.; Saunders, Jane (1963). "Liberación de organismos de textiles contaminados" . The Journal of Hygiene . 61 (4): 507– 513. doi : 10.1017/s0022172400021136 . JSTOR 3860927. PMC 2134562. PMID 14099028 .
- ↑ Simard, Carole; Trudel, Michel; Paquette, Gilles; Payment, Pierre (1983). "Investigación microbiana del aire en un edificio de apartamentos" . The Journal of Hygiene . 91 (2): 277– 286. doi : 10.1017/S0022172400060290 . JSTOR 3862904. PMC 2129367. PMID 6358346 .
- ↑ Sorber, Charles A.; Bausum, Howard T.; Schaub, Stephen A.; Small, Mitchell J. (1976). "Un estudio de aerosoles bacterianos en un sitio de riego de aguas residuales". Journal (Water Pollution Control Federation) . 48 (10): 2367– 2379. JSTOR 25040032. PMID 792471 .
- ↑ Weis, Christopher P. (2002). "Aerosolización secundaria de esporas viables de <EMPH TYPE="ITAL">Bacillus anthracis</EMPH> en una oficina contaminada del Senado de los Estados Unidos". JAMA . 288 (22): 2853– 2858. doi : 10.1001/jama.288.22.2853 . PMID 12472327 .
- ↑ Spear, Terry M.; Svee, Wayne; Vincent, James H.; Stanisich, Nick (1998). "Especiación química del polvo de plomo asociado con la fundición primaria de plomo" . Environmental Health Perspectives . 106 (9): 565– 571. doi : 10.2307/3434231 . JSTOR 3434231. PMC 1533158. PMID 9721256 .
- ↑ "El 'muestreador' de aire en Dugway ayuda a los estudios" . The Salt Lake Tribune . 24 de enero de 1958 – vía newspaperarchive.com.
- ↑ "Un residente de Provo desarrolla un instrumento para estudiar la contaminación del aire" . Utah Provo Daily Herald . 26 de enero de 1958 – vía newspaperarchive.com.
- ↑ "Hacen que la contaminación pague" . The Deseret News . 7 de noviembre de 1970 – vía Google News Archive.
- ↑ Mitchell, JP; Costa, PA; Waters, S. (1988). "Una evaluación de un impactador en cascada Andersen Mark II". Journal of Aerosol Science . 19 (2): 213– 221. Bibcode : 1988JAerS..19..213M . doi : 10.1016/0021-8502(88)90224-8 .
- ↑ King, Maria D.; McFarland, Andrew R. (2012). "Uso de un muestreador de bioaerosoles Andersen para proporcionar simultáneamente distribuciones de tamaño de partículas cultivables y organismos cultivables". Aerosol Science and Technology . 46 (8): 852– 861. Bibcode : 2012AerST..46..852K . doi : 10.1080/02786826.2012.669507 .
- ^ Xu, Zhenqiang; Wei, Kai; Wu, Yan; Shen, Fangxia; Chen, Qi; Li, Mingzhen; Yao, Maosheng (2013). "Mejora del muestreo de bioaerosoles mediante impactadores Andersen utilizando una placa de agar esparcida con aceite mineral" . MÁS UNO . 8 (2) e56896. Código Bib : 2013PLoSO...856896X . doi : 10.1371/journal.pone.0056896 . PMC 3584084 . PMID 23460818 .
Lecturas adicionales
- Gillespie, VL; Clark, CS; Bjornson, HS; Samuels, SJ; Holland, JW (1981). "Una comparación de impactadores Andersen de dos y seis etapas para aerosoles viables". American Industrial Hygiene Association Journal . 42 (12): 858– 864. doi : 10.1080/15298668191420819 .
- Blomquist, Göran; Palmgren, Urban; Ström, Gunnar (1984). "Técnicas mejoradas para el muestreo de partículas fúngicas en el aire en entornos altamente contaminados" . Scandinavian Journal of Work, Environment & Health . 10 (4): 253– 258. doi : 10.5271/sjweh.2334 . JSTOR 40965745. PMID 6494845 .
- Solomon, William R. (1970). "Una aplicación simplificada del muestreador de Andersen para el estudio de partículas de hongos en el aire". Journal of Allergy . 45 (1): 1– 13. doi : 10.1016/0021-8707(70)90012-2 . PMID 5262360 .
- Lednicky, John A.; Fennelly, Kevin; Tribby, Matthew; Wu, Chang-Yu; Heil, Gary L.; Radonovich, Lewis; Loeb, Julia (2014). "Recolección y medición de aerosoles de virus de influenza viables en medios líquidos en un impactador en cascada Andersen" . Virus Adaptation and Treatment : 1. doi : 10.2147/VAAT.S74789 .
- Greene, VW; Vesley, D. (1962). "Método para evaluar la eficacia de las mascarillas quirúrgicas" . Journal of Bacteriology . 83 (3): 663– 667. doi : 10.1128/jb.83.3.663-667.1962 . PMC 279325. PMID 13901536 .
- Seguridad y salud en el trabajo
- Medición de aerosoles